1.CE certificate 2.ISO 13485 3.Series Hemodialysis Equipment Equipments(HDF-online)
sws-4000B Hemodialysis Equipment is applied to HD, IUF-bibag, IUF, HP etc., substitution fluid weighing device enables accurate control of fluid balance.
SWS-4000 Hemodialysis Equipment is applied to HD, HF-online, IUF, HP etc, produce substitution fluid on -line.
SWS-4000 Hemodialysis Equipment
Forms of Therapy:
Hemodialysis, Online Hemofiltration (Online HF), Online Hemodiafiltration (Online HDF),Isolated
Ultrafiltration, Sequential Ultrafiltration, Hemoperfusion, Plasma exchange (optional).
Technical Features:
1. The Multi-Functional Machine Offers Customized Curing Mode For Patients.
2. The Advanced Volume Balance Control System Enables Accurate Control of Ultrafiltration.
3.. Double-Stage Intoxication Filtration Ensuring That The Substitution Fluid Meets AAMI Standards.
4. Reliable System and Safe Treatment Ensured By Pressure Detectors, Conductivity Detectors, Air Detectors and Blood Leakage Detectors.
Function Features:
1. Adjustable sodium And UF profiling Enable Customized Treatment For Patients.
2. Convenient Process for Sequential Ultrafiltration and Dialysis.
3. High Volume Hemofiltration (HVHF) Is Available
4. Kt/V Calculation Function Can Be Used For Evaluating the Sufficiency of Dialysis via the Urea Kinetic Model Volume Balance System.
5. Isolated Hemoperfusion Mode, Decrease the Normal Hemodiafiltration Device Self-Detecting Time And The Dialysate Waste.
6. The Adaptive Regulation and Control System Effectively Avoids Various Complications. And Can Clear Middle and Small Molecules Toxin.
7. Automatic Disinfestation and Decalcifilcation by Many Kinds of Disinfectant Liquids and Picking Solutions.
8. Back-Up Power Supply For 30 Min After Electricity Off.
Technical Data:
DIMENSIONS (W x H x L) |
930 × 854×1710 MM |
Gross Weight |
Aprox. 140 Kg |
Power Supply |
AC220V, 50~60HZ |
Connected Load |
≤2000VA |
Water Supply |
Pressure Range :0~8.0BAR, Temperature
Range:0~35°C |
Venous Pressure Working Range |
-700~+800mmhg, Accuracy±10mmhg |
Arterial Pressure Working Range |
-700~+800mmhg, Accuracy±10mmhg |
Transmembrane Pressure Working
Range |
-700~+800mmhg, Accuracy±10mmhg |
Blood Flow Range |
0~700ml/Min |
Dialysate Flow Range |
0~300~500~800ml/Min (Adjustable) |
Substitution Fluid Flow |
0~25L/H |
Dialysate Temperature Working Range |
30~41°C,Accuracy±0.1°C |
Heparin Pump Working Range |
0~10 ml/H,Accuracy±0.1ml/H |
Conductivity Working Range |
10~20 ms/cm Accuracy±0.1ms/cm |
Air Detector |
Detecting≥0.02ML/Single Bubble Detecting |
Blood Leak Detector |
Detecting blood leak ≥0.35mL/min (HCT 32%) |
Hot Rinse |
≥85°C |
Clean/ Disinfection | Decalcification: Rinse, citric acid Thermal Disinfection: Hot water or citric acid Chemical Disinfection: Sodium hypochlorite or peracetic acid |