infant radiant warmer 1.Tilt angle of infant bed is adjustable 2.Transparent protector folded 3.Temperature deviation
EMS-100 (Standard) Infant Radiant Warmer
Brief introduction:
1. Microprocessor based servo controlled temperature system.
2. The microprocessor-controlled unit allows for three settings: Pre-warm, manual or automatic.
3. Alarm functions for safety includes: Power failure, temperature deviation, Temperature sensor failure, over temperature.
4. APGAR timer to record therapy time
5. Convenient X-ray cassette under infant bed
6. Heater head can be adjusted +90 degree horizontally
7. Tilt angle of infant bed is adjustable
8. Observe light angle is adjustable
9. Transparent protector folded
10. Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source, the heating is quick and life-span is long
11. Alarms are audible and visual
12. A second thermal cut-out function for more safety
13. Skin temperature sensor failure protect function avoid over temperature
14. Temperature deviation can be amended in front panel directly
Standard configuration
Main body (including the radiant source, control system, infant bed, bracket), X-ray cassette, Observe light, I. V. Pole, skin temperature sensor, tray, mattress, transparent protector, castors.
Optional parts: Drawer, RS-232(485) connecter
Technical data:
1. Power supply: 110/220V 60/50Hz
2. Power input: ≤ 1200VA
3. Operating condition
An ambient Temperature: 18°C-30°C
An ambient Relative Humidity: 30--75 percent RH
Atmospheric Pressure: 700~1060 hPa
An ambient Air Velocity <0.3m/s
4. Temperature control: 32~38°C
5. Temperature control accuracy: ≤ 0.5°C
6. Mattress temperature uniformity: <25°C
7. Skin temperature sensor accuracy: +/- 0.3°C
8. Warm-up time: ≤ 30min
9. Infant bed tilt angle: + 10degree;
10. APGAR timer range: ≤ 24h
11. Heater head horizontal: 90degree;
12. Mattress size: 67cm X 54cm
13. Distance from heater to mattress: 80cm
Transport and storage
Environment Temperature: -40~ +55°C
Environment Relative Humidity: ≤ 95percert Atmospheric Pressure: 500~1060hPa
Package:
Each unit is packed one case
Size: 120cmX 70cm X 96cm
Gross weight: ≤ 90kg
EMS-100 (Standard) Infant Radiant Warmer