TH-2008 Calibration Dynamic Gas Generator
TH-2008 Calibration Dynamic Gas Generator

Using dilution technology. Supply of precise levels of carbon monixide, sulfur dioxide,and nitrogen monoxide.

TH-2008 Calibration Dynamic Gas Generator is used to calibrate various ambient air analyzers.

It is used together with the Zero Generator and various standard gas cylinders. Using Zero gas dilution standard gas method, the TH-2008 can complete the calibration to analyzer through presupposed programming or user's random operation. The calibration includes zero calibration, span calibration, multi-point calibration and efficiency detection in reduction room in which NOx is analysed.

Main Technical Parameter

Diluting gas inlet number1
Source gas inlet number3
Diluting mass flow controller0-10L/min(extensible)
Standard gas mass flow controller0-50ml/min(extensible)
Flow rate accuracy±1.0%
Flow rate repetition0.15%F.S
Linearity±0.15%F.S
Inlet pressure100kpa-200kpa
Response time(T95)≤60s
Diluting rate200:1-2000:1(standard)
Outlet number3
Operation temperature0°C-40°C
 Ozone calibration device optional
 Meteorological titration device optional