ECG Machine Lead standard:12 leads Frequency response:150Hz Rechargeable battery supporting 50 patients examinations ecg
ECG MACHINE LDM-901A
Features: Single channel digital ECG machine with measurement report Simultaneous acquisition of 12-lead ECG signal 8 patients save/copy/communication 320X240 graphic, 3.8 inch screen large LCD display 12-lead ECG waveform Paper size: 50mmx20/30m Built-in RS232 & USB interface AC/DC power, built-in rechargeable Ni-MH battery This device have Analysis Result: MI & AF 881 Atrial Fibrillation 882 Atrial Flutter 883 Atrial Fibrillation(Slow) 884 Atrial Fibrillation(Rapid)SOFTWARE CD is option with USB Key:
Specification:
Safety Mode | IEC60601-1 class I, type CF |
Lead | Standard 12 leads |
Lead acquisition | Simultaneously 12 leads |
Resolution | 18bit/1000Hz |
Record mode | Manual mode, 3 Auto modes(with measurement report) |
Filter | AC, EMG filter |
Drift filter | Anti-drift system |
CMRR | >60dB, >100dB(with AC filter) |
Input circuit | Floating, Protection circuit against Defibrillator effect |
Input circuit current | =<0.1μA |
Input Impedance | >=50MΩ |
Patient current leakage | < 10μA |
Calibrating voltage | 1mV+_2% |
Voltage Tolerance | +_500mV |
Time constant | >3.2s |
Frequency response | 0.05 ~150Hz |
Noise level | =<15μVp-p |
Sensitivity | =<20μV |
Inter-channel interference | =<0.5mm |
Paper speed | 25, 50mm/s(+_3%) |
Sensitivity | Auto, 2.5, 5, 10, 20mm/mV |
Recorder | High resolution thermal printer |
Recording mode | Thermal Array printing |
Paper specification | 50mmx20/30m, roll |
LCD display | 320x240 graphic LCD, display status&waveform |
Power supply | AC: 110V/220V/85~265V, 50/60Hz+_1Hz, 30VA |
DC: 12V(1500mAh) rechargeable Ni-MH battery | |
Dimension | 288x204x60mm |
Packing | 380x300x210mm |
Weight | 4.3kg(gross), 3.27kg(net) |
Digital Single Channel ECG Machine /Monitor Electrocardiograph