400mA Medical X-ray machine YSX0115 1 400mA medical x-ray machine 2 medical x-ray equipments 3 medical x-ray unit
Medical X-ray machine YSX0115
1 Equipped with single diagnostic table and x-ray rubes, single direct bridge full-wave commutation, tube voltage is 125kV, tube currency is 400mA.
2 X-ray tube capacity protection, anode starting-up protection and radiograph time-limit protection device semi-conduct controlled, fluoroscopy computed time device, x-ray tube over-hot protectorate provided, ensure x-ray tube run safely long time, improve accuracy of kV mA S.
3 Bigger scope of rotation diagnostic table, movable table muli-item spot-film device, easy to operate, radiation –proof, ensure operate and the tested.4 Radiography table can move lengthways and crossover, can loaded all kinds of radiography dark box, use conveniently and it is beautiful.5 Applied in all kinds of hospital for radiography such as general body,beam filtration,and chest.
Department | Technology parameters | ||||
High voltage generator control table and high voltage transform assemble | Fluoroscopy | Tube currency | 0.3-5mA continuous currency | ||
Tube voltage | 45-110KV continuous voltage | ||||
Radiography | Tube currency | 50-400mA shift | |||
Tube voltage | 45-125KV continuous voltage | ||||
Exposure time | 0.03-5s,23 shift | ||||
Power | Voltage | 380V±10% | |||
Frequency | 50Hz±1Hz | ||||
Capacity | ≥30KVA | ||||
Inner resistance | ≤0.3Ω | ||||
X-ray assembly | Type of x-ray | XD51-20.40/125 | |||
Focus | 1×1.2×2 double focal | ||||
Rated rotate speed of anode | 2800r/min | ||||
CDGI Multi function radiography table | Moving table | Lengthways | ±600mm | Electromagnetism locking | |
Across | ±120mm | ||||
Mobile beam filtration grid | Grid density | 28L/cm | |||
Grid ratio | 8:01 | ||||
Focusing distance | 100cm | ||||
Scope of moving | ≥600mm | ||||
Radiography item | 5”×7”,8”×10”,10”×12”,11”×14”, | ||||
12”×15”,14”×14”,14”×17” | |||||
Distance from surface to film | 75mm | ||||
Height of bed | 700mm | ||||
Moving pillar | ±180°,per 90°orientation, moving lengthways ≥1500mm | ||||
Bulb tube | Rotate the cross arm±180,ascend $ descend 1100mm | ||||
Option |
| ||||
LSJ pillar radiography frame | The scope of ascend and descend | 1450mm | |||
The minimum distance from focal spot | 380mm | ||||
Radiography item | 8”×10” 10”×12” 11”×14” 12”×15”14”×14” 14”×17” | ||||
Fixed beam filtration role | Grid density | 40L/cm | |||
Grid radio | 10:01 | ||||
Focal distance | 180cm |
400mA Medical X-ray machine YSX0115